Simulation results of a small animal liquid xenon PET detector

Résumé

Monte Carlo simulations of a novel concept PET detector for small animal imaging are presented. The scintillation medium of the detector is liquid xenon whose characteristics in terms of detection rival with the common scintillator crystals. Moreover, the axial geometry of the detector enables depth of interaction measurement. A detector module has been built and an experimental test bench has been developed. Simulations of the test bench enabled to determine the methods to use for analysing the experimental data. Moreover, they indicate the spatial resolution in the axial direction and the energy resolution which can be expected from the detector. The results show an axial resolution of 2.87plusmn0.12 mm and an energy resolution of 7.59plusmn0.34%.

Mots clés

Fichier non déposé

Dates et versions

hal-00308858 , version 1 (02-08-2008)

Identifiants

Citer

Yannick Grondin, Michel Desvignes, Laurent Desbat, Stéphane Mancini, M.L. Gallin-Martel, et al.. Simulation results of a small animal liquid xenon PET detector. ISBI 2008 - 5th IEEE International Symposium on Biomedical Imaging, May 2008, Paris, France. pp.1227-1230, ⟨10.1109/ISBI.2008.4541224⟩. ⟨hal-00308858⟩