Evaluation of the Needle Positioning Accuracy of a Light Puncture Robot Under MRI Guidance: Results of a Clinical Trial on Healthy Volunteers - Archive ouverte HAL Accéder directement au contenu
Article Dans Une Revue CardioVascular and Interventional Radiology Année : 2018

Evaluation of the Needle Positioning Accuracy of a Light Puncture Robot Under MRI Guidance: Results of a Clinical Trial on Healthy Volunteers

Résumé

PURPOSE: To assess the accuracy of Light Puncture Robot (LPR) as a patient-mounted robot, in positioning a sham needle under MRI guidance for abdominal percutaneous interventions. MATERIALS AND METHODS: This monocentric, prospective and non-controlled study was approved by the ethics review board. The study evaluated the accuracy of LPR V3 to achieve a virtual puncture in 20 healthy volunteers. Three trajectories were tried on each volunteer, under 3-T MRI guidance. RESULTS: Accuracy under 5 mm in attaining a 10 cm-deep target was reached in 72% of attempts after 2 robot motions with a median error of 4.1 mm [2.1; 5.1]. Median procedure time for one trajectory was 12.9 min [10.2; 18.0] and median installation time was 9.0 min [6.0; 13.0]. CONCLUSION: LPR accuracy in the deployment of a sham needle inside the MRI tunnel and its setup time are promising. Further studies need to be conducted to confirm these results before clinical trials.
Fichier non déposé

Dates et versions

hal-02264142 , version 1 (06-08-2019)

Identifiants

Citer

Julien Ghelfi, Alexandre Moreau-Gaudry, Nikolai Hungr, Céline Fouard, Baptiste Veron, et al.. Evaluation of the Needle Positioning Accuracy of a Light Puncture Robot Under MRI Guidance: Results of a Clinical Trial on Healthy Volunteers. CardioVascular and Interventional Radiology, 2018, 41 (9), pp.1428-1435. ⟨10.1007/s00270-018-2001-5⟩. ⟨hal-02264142⟩
2032 Consultations
0 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More